Absorbed dose
energy deposited per unit mass
Gray = J /kg
KERMA
Kinetic Energy released per unit mass
Gray
Foetal dose-termination of pregnancy
1. < 1000 mGy : no termination
2. 100-500 mGy : optional
3. > 500 mGy : recommend
X-tube V and A
tube voltage = kV = accelerating voltage
tube current = mA
good anode
-high atomic number
-good isolation
-high melting point of target material
-low vapour point
contrast resolution
contrast increase
- structure thickness increase
- kV deceases
spatial resolution
line pairs per mm
quantum noise
- reduces contrast resolution.
- signal-to-noise
SNR = N/squr {N}
ratio of scatter
scattered : primary radiation
4 : 1 =thorax
9 : 1 = pelvis
-Pt thickness
-field size
how to reduce scatter?
-decrease field size
-decrease pt thickness
-decrease kV
-use a grid
-increase object-image distance
DR
indirect versus direct
indirect:
- more x-ray absoprtion
- lower dose
- less noise
direct:
- better spatial resolution
- faster, no cassette
mammography
- lower kV
- lower Z (atomic number)
- 1.3 - 3 mGy per exposure